ClearanceBrief
Recall Z-2799-2016 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Siemens Healthcare Diagnostics Inc

Siemens RAPIDPoint 500 Blood Gas Analyzer Siemens Material Number (SMN): 10492730, 10696855, 10696857, 10697306

DateAug 11, 2016
StatusTerminated
ReasonThere is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
Root causeSoftware design
ActionSiemens Healthcare issued Urgent Field Safety Notice to (US) customers via Federal Express. . All affected customers outside the US provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs customers of the issue, Actions to be Taken by the Customer and to return the response form. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative
Quantity6786 units
DistributionNationwide Foreign: Albania Algeria Angola Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bosnia Herzeg. Botswana Brazil Brunei Bulgaria Canada Chile Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Hong Kong H
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113216RAPIDPOINT 500 BLOOD GAS ANALYER
Siemens Healthcare Diagnostics, May 3, 2012