Recall Z-2732-2016 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System. Catalog No. 629029, B24465, B24802, B68304 (Unicel DxH 800 Coulter Cellular Analysis System). Catalog No. B23858 (Unicel DxH600 Coulter Cellular Analysis System).

DateSep 2, 2016 (initiated Aug 8, 2016)
StatusTerminated
ReasonThere is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.
Root causeSoftware design
ActionAn Urgent Medical Device Recall letter dated 8/8/16 was sent to all customers who purchased the Unicel DxH 800 and DxH 600 Coulter Cellular Analysis System to inform them that there is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support; 800-526-7694 in t
Quantity5,282 units total (2,877 units in US)
DistributionAffected geographies: Andorra, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Columbia, Croatia, Czech Republic, Egypt, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Le
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140911UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM/ UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM