Recall Z-2735-2016 ยท KPS System, Tomography, Computed, Emission
Philips Medical Systems (Cleveland) Inc
BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging
| Date | May 25, 2016 |
| Status | Terminated |
| Reason | The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script. |
| Root cause | Software design |
| Action | The firm is informing the consignee of this issue through Customer Information Letter. |
| Quantity | 117 units |
| Distribution | Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, N |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K062298 | APOLLO GAMMA CAMERA SYSTEM Adac Laboratories, Aug 23, 2006 |