Recall Z-2735-2016 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging

DateMay 25, 2016
StatusTerminated
ReasonThe firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
Root causeSoftware design
ActionThe firm is informing the consignee of this issue through Customer Information Letter.
Quantity117 units
DistributionWorldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, N