Recall Z-2848-2016 ยท OMP Negative Pressure Wound Therapy Powered Suction Pump

Smith & Nephew, Inc.

RENASYS Sterile Transparent Film and all the RENASYS Dressing kits that contains the RENASYS Sterile Transparent Film Is intended for patients who would benefit from a suction device (negative pressure) to help promote wound healing by removing fluids including irrigation and body fluids, wound exudates and infectious materials. Examples of appropriate wounds types include chronic, acute, traumati

DateJul 21, 2016
StatusOpen, Classified
ReasonSterility seal integrity failures on the Renasys Sterile Transparent Film
Root causeUnknown/Undetermined by firm
ActionSmith & Nephew sent an Urgent Medical Device Removal Notice dated July 21, 2016, to all consignees informing them of a voluntary medical device recall involving Sterile RENASYS Transparent Film and the RENASYS Dressing Kits that contain the folded RENASYS Transparent Film. Smith & Nephew have identified that the process of folding the Sterile RENASYS" Transparent Film pouch during its packaging has the potential to damage the pouch seals resulting in a loss of sterility of the product. The loss of sterility has the potential to contribute to a local infection. Actions to be taken by cust
Quantity45,455 devices (in the USA)
DistributionNationwide Distribution to AK, AL, AR, AZ, CA, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TX, UT, VA, WA, WI, Hawaii, and Puerto Rico Argentina, Australia, Bermuda, Brazil, Brunai, Canada, Curacao, Germany, Great Britain, Hong Kong, Ireland, Jordan, M
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061919VERSATILE 1 EZCARE WOUND VACUUM SYSTEM
Blue Sky Medical Group Incorporated, Feb 5, 2007
K142979RENASYS; EZ MAX Negative Pressure Wound Therapy; Foam NPWT Wound Dressing Kits with Soft Port; Gauze NPWT Wound Dressing
Smith & Nephew, Inc., Apr 29, 2015