ClearanceBrief
Recall Z-2545-2016 ยท MGZ Valvulotome

LeMaitre Vascular, Inc.

1.5 mm HYDRO LeMaitre Valvulotome, 40 cm. Sterile REF # 1010-00 Product Usage: The HYDRO LeMaitre Valvulotome is intended for the treatment of vascular disorders and more particularly for excising or disrupting venous valves

DateAug 3, 2016
StatusTerminated
ReasonWire which deploys the loop may be come detached at the handle causing the device to be inoperable
Root causeProcess control
ActionLeMaitre Vascular, Inc. sent an Urgent Medical Device Recall letter dated June 29, 2016 to customers via priority mail. The letter identified the affected product, problem and actions to be taken. Customers were requested to locate, quarantine and return the devices. Customers were instructed to complete and return the response form to LeMaitre Vascular, Inc. as a record of notification and reconciliation. Any unused product will be requested to be returned and replaced. For questions call 781-221-2266 ext. 183.
Quantity50 units
DistributionWorldwide Distribution - US Nationwide and in the countries of Australia, BELGIUM, BRAZIL, CANADA, Czech Republic, DENMARK, FINLAND, FRANCE, GERMANY, Italy, Japan, NETHERLANDS, Russia, Spain, SWEDEN, Switzerland, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1400421.5MM HYDRO EXPANDABLE LEMAITRE VALVULOTOME
LeMaitre Vascular, Inc., Apr 10, 2014