Recall Z-2626-2016 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system

DateAug 23, 2016 (initiated Jul 13, 2016)
StatusTerminated
ReasonDue to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
Root causeSoftware in the Use Environment
ActionSiemens mailed to customers an Important Customer Safety Notice to inform about a possible malfunction when using the affected systems and what actions to take to eliminate the problem. Also, Siemens informed their customers that they were currently developing a new firmware to eliminate the internal communication error, which should be released in August 2016. Customers were asked to contact their service organization at 800-888-7436, if they had any questions.
Quantity4 systems
DistributionDistributed to: MI, NY, CA, KY, ND, NE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143401SOMATOM Definition Edge
Siemens Medical Solutions USA, Inc., Apr 6, 2015