Recall Z-2663-2016 ยท JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use

Tecan US, Inc.

Infinite 200 PRO The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conf

DateAug 25, 2016 (initiated Jul 25, 2016)
StatusTerminated
ReasonThe batteries have the potential to overheating, melt or char.
Root causeNonconforming Material/Component
ActionTecan contacted all customers by Safety Announcement, via certified mail starting July 25, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Because affected batteries pose a fire and burn hazard, it is extremely important that customers check whether their battery is affected. Customers were advised to go immediately to the HP Battery Recall website at: http://www.HP.com/go/batteryprogram2016 to check if their battery is affected by this recall. Customers were also advised to complete and return the Customer Response Form at QA@tecan.com or
Quantity27 units
DistributionWorldwide Distribution - US including AK, AL, AR, CA, FL, GA, HI, KY, MA, MD, NC, NY, PA, PR, SC, TN, TX, UT, VA. International account: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.