Recall Z-0314-2017 ยท GHH Fibrin Split Products

BBI SOLUTIONS OEM LTD

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

DateJun 1, 2016
StatusTerminated
ReasonA component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.
Root causeNonconforming Material/Component
ActionBBI Solutions sent an Urgent Medical Device Recall letter dated June 30, 2016, to all affected customers. Customers were advised to immediately discontinue use of the product, consider the requirement for review of reported test results, complete and return the customer acknowledgement form to BBI Solutions and to complete and return the customer reconciliation form to the distributor. Customers who have forwarded product to other sites, should inform those customers of the product recall and provide them with a copy of the notice. Customers with questions were instructed to call Tel: +44
Quantity420 units (183 units quarantined)
DistributionUS distribution to CT and IN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K905643BED RED D-DIMER WHOLE BLOOD AGGLUTINATION ASSAY
Agen Biomedical , Ltd., Apr 12, 1991