Recall Z-2363-2016 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.

DateFeb 5, 2016
StatusTerminated
ReasonWhen measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
Root causeSoftware design
ActionThe recalling firm's customer support notified customers via phone call beginning 2/5/2016 to inform them of the issue. Customers were informed that a software fix that corrects the issue was available. Customers were asked not to use the measurements on mammo images until upgraded. If customers want an upgrade, then someone would contact them to schedule the upgrade.
Quantity49 sites potentially have the affected versions
DistributionWorldwide Distribution -- USA and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082144AMICAS PACS, VERSION 6.0
Amicas, Inc., Sep 5, 2008