Recall Z-0313-2017 ยท MJX Kit, Serological, Positive Control

Ortho-Clinical Diagnostics

VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro qualitative detection of Hepatitis B Surface Antigen (HBsAg) in human serum and plasma (heparin, EDTA o

DateJun 22, 2016
StatusTerminated
ReasonThe signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.
Root causeUnder Investigation by firm
ActionThe firm, Ortho Clinical Diagnostics, distributed "Important Product Correction Notification" letters (Ref. CL2016-129) dated 6/22/2016 and Confirmation of Receipts to their customers via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only). Foreign affiliates were informed by email (on 22 June 2016) of the issue and instructed to notify their consignees of the issue and required actions. All customers who received the affected product were instructed to immediately discontinue using and discard the affected lots. Forward a
Quantity1,621 units
DistributionWorldwide Distribution-USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030067VITROS IMMUNODIAGNOSTIC PRODUCTS HBSAG CONTROLS
Ortho-Clinical Diagnostics, Jan 17, 2003