Recall Z-2630-2016 ยท JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered

Stryker Instruments Div. of Stryker Corporation

120 V Neptune 3 Rover, Model Number: 0703-001-000

DateAug 23, 2016 (initiated Jul 1, 2016)
StatusTerminated
ReasonStryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.
Root causeUnder Investigation by firm
ActionStryker Instruments initiated a voluntary recall of the Stryker Neptune 3 Rover via certified mail on 07/01/2016, due to the potential for the Neptune 3 Rover to experience system errors if there are high levels of electromagnetic interference in the operating room. 1. Immediately review the Recall Notification. 2. Confirm that the Rover serial numbers identified on the Business Reply Form (BRF) are at your facility. 3. A Stryker representative will facilitate the repair of your affected Neptune 3 Rover(s). Please coordinate the availability of the product to assure the repair can be im
Quantity78
DistributionDomestic:MI, CA, ID, IA VA/DOD: None Foreign: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153407Stryker Neptune 3 Waste Management System
Stryker Corporation, Apr 25, 2016