Recall Z-2689-2016 ยท BSZ Gas-Machine, Anesthesia

Mindray DS USA, Inc. dba Mindray North America

A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international)

DateJun 15, 2016
StatusTerminated
ReasonPotential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System.
Root causeNonconforming Material/Component
ActionMindray sent via certified mail with return receipt a recall letter dated June 15, 2016 to their affected customers. Mindray will replace the gasket on all e-size cylinder yokes on the affected A7 systems. Customer was instructed to contact their local Mindray Service Representative to arrange for this replacement. Customers can continue to use their A7 system while awaiting the replacement of the gasket(s). Customers can contact Ms. Diane Arpino, Director, Quality Operations and Regulatory Affairs via email to: d.arpino@mindray.com or via telephone to: (201)995-8407
Quantity167 units (165 units - US) and (2 units - International)
DistributionUS Nationwide Distribution to AL, MA, MD, MN, MO, NE, NJ, OK, OR, PA and VA; and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151954A7 Anesthesia System
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Oct 27, 2015