ClearanceBrief
Recall Z-2569-2016 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Medtronic

Valleylab Laparoscopic Curved Spatula Electrode 28CM Item Code: E278128 The Valleylab Laparoscopic Handset and Electrodes are indicated for use in laparoscopic and thoracoscopic surgical procedures where the delivery of electrosurgical current, irrigation and suction via a single handset is desirable

DateJul 8, 2016
StatusTerminated
ReasonPotential for compromise of the package resulting in a breach of the sterile barrier
Root causeProcess control
ActionMedtronic/Covidien sent an Urgent Medical Device Recall letter dated July 8, 2016, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Customers with questions were instructed to contact their Medtronic representative or Customer Service at 800-882-5878.
DistributionWorldwide Distribution - US (nationwide) and Internationally to Australia, Canada, Japan, Korea, Hong Kong, Thailand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K926481VALLEYLAB MODEL E252OH, FORCE GSU(TM) HANDSET
Valleylab, Inc., Feb 12, 1994