Recall Z-2306-2016 ยท IXR Table, Radiographic, Tilting

Philips Electronics North America Corporation

UroDiagnost; x-ray system. Product Code: 708033

DateJul 19, 2016
StatusTerminated
ReasonThe two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag
Root causeDevice Design
ActionOn July 19, 2016, Philips sent a Field Safety Notice to all affected customers. Philips Field Service Engineers will add an independent safety catch free of charge.
Quantity993 total
DistributionWorldwide distribution. US nationwide, Canada, Algeria, ARGENTINA, Australia, Austria, Belgium, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Germany, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Republic of Korea