Recall Z-2305-2016 ยท IZI System, X-Ray, Angiographic

Philips Electronics North America Corporation

MultiDiagnost Eleva with Flat Detector; x-ray system. Product Codes: 708034, 708037, 708038.

DateJul 19, 2016
StatusTerminated
ReasonThe two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag
Root causeDevice Design
ActionOn July 19, 2016, Philips sent a Field Safety Notice to all affected customers. Philips Field Service Engineers will add an independent safety catch free of charge.
Quantity993 total
DistributionWorldwide distribution. US nationwide, Canada, Algeria, ARGENTINA, Australia, Austria, Belgium, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Germany, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Republic of Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023441PHILIPS MULTIDIAGNOST ELEVA
Philips Medical Systems, Oct 30, 2002