Recall Z-2389-2016 ยท HRS Plate, Fixation, Bone

Synthes (USA) Products LLC

Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the following systems: - 2.4. mm Titanium LCP Distal Radius Plate System - Titanium LCP Compact Distal Radius Sys

DateAug 11, 2016 (initiated Jun 28, 2016)
StatusTerminated
ReasonIt was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.
Root causeNonconforming Material/Component
ActionThe firm, DePuy Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated July 14, 2016 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to review your inventory and immediately remove and return all unpackaged Locking Screws Self-Tapping with Stardrive Recess 12mm in any affected kits along with any screws still in the original packaging with the part number and lot number identified in the recall notification. Customers were also instructed to call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorizati
Quantity41 units
DistributionUS Distribution to states of: AR, CA, CO, DE, FL, GA, IA, IL, KY, LA, MN, NJ, OK, PA, SD, TN, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012114SYNTHES LOCKING DISTAL RADIUS PLATING SYSTEM
Synthes (Usa), Sep 28, 2001
K050110SYNTHES (USA) LCP MODULAR FOOT PLATES
Synthes (Usa), Mar 4, 2005
K063049SYNTHES (USA) MODULAR MINI FRAGMENT LCP SYSTEM
Synthes (Usa), Dec 18, 2006
K1026942.4MM VA-LCP DORSAL DISTAL RADIUS PLATES
Synthes (Usa), Dec 9, 2010
K1032432.4MM VA-LCP INTERCARPAL FUSION SYSTEM
Synthes (Usa), Jan 21, 2011