Recall Z-2395-2016 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

ARTIS Q AND Q.ZEN- MODULAR ANGIOGRAPHIC SYSTEM, interventional fluoroscopic x-ray system22 AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis, Artis zee /

DateJul 13, 2016
StatusTerminated
ReasonThe possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Root causeUnder Investigation by firm
ActionSiemens has implemented a Customer Safety advisory Notice to all affected customer via Update Instruction AX007/15/S. Additional, a corrective action will be implemented via Update Instructions AX006/15/S and AX003/16/S.
Quantity11,111, worldwide and 3,187 in the USA
DistributionNationwide Distribution to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WV,WY, Puerto Rico and Guam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123529ARTIS Q AND Q.ZEN- MODULAR ANGIOGRAPHIC SYSTEM
Siemens Medical Solutions USA, Inc., Feb 26, 2013