Recall Z-2392-2016 ยท JAK System, X-Ray, Tomography, Computed

Hitachi Medical Systems America Inc

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

DateJun 30, 2015
StatusTerminated
ReasonThe firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Root causeComponent design/selection
ActionHitachi sent an Device Correction Notice dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. Customers were informed that there is a problem with the Hitachi Medical Systems America, Inc.CT system; the issue is with a power supply in the CT gantry sub-system. During fast rotation (i.e., 35 sec), the centrifugal force applied to the power supply may cause it to fail over time. For further questions, please call (330) 425-1313 ext. 3720.
Quantity33 units
DistributionUS Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101888SCENARIA WHOLE-BODY X-RAY CT SYSTEM
Hitachi Medical Systems America, Inc., Apr 8, 2011