Recall Z-2350-2016 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
CURVED-TIP STAPLER 30,IS4000, model number 470530; Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pedia
| Date | Aug 5, 2016 (initiated Jul 14, 2016) |
| Status | Terminated |
| Reason | An increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend |
| Root cause | Software design |
| Action | Field Safety Notifications - Urgent Medical Device Corrections and Removal letters were sent on July 15, 2016 (dated July 14, 2016 advised users of the Stapler Firing Failed message and actions to take should the message be received. |
| Quantity | 2065 total |
| Distribution | US, including Puerto Rico and Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Monaco, Netherlands, Norway, Portugal, , Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Venezuela. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K152421 | IS4000 Stapler 30 Instrument and Stapler 30 Reloads Intuitive Surgical, Inc., Mar 4, 2016 |