ClearanceBrief
Recall Z-2348-2016 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

STAPLER 45, IS4000, model number 470298 Gastroenterology/Urology: The Intuitive Surgical da Vinci EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line

DateAug 5, 2016 (initiated Jul 14, 2016)
StatusTerminated
ReasonAn increase in the number of da Vinci EndoWrist Stapler complaints where the customer experienced a Stapler Firing Failed message as well as review of da Vinci Xi Surgical System logs found an increased trend
Root causeSoftware design
ActionField Safety Notifications - Urgent Medical Device Corrections and Removal letters were sent on July 15, 2016 (dated July 14, 2016 advised users of the Stapler Firing Failed message and actions to take should the message be received.
Quantity2065 total
DistributionUS, including Puerto Rico and Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Monaco, Netherlands, Norway, Portugal, , Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140553ENDOWRIST STAPLER 45 AND STAPLER 45 RELOADS
Intuitive Surgical, Inc., Jul 25, 2014