Recall Z-0296-2017 ยท GDW Staple, Implantable

Ethicon Endo-Surgery Inc

Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses

DateOct 20, 2016 (initiated Jul 15, 2016)
StatusTerminated
ReasonQuality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.
Root causeDevice Design
ActionEthicon sent an Urgent Medical Device Recall letter dated July 15, 2016, to all affected consignees via UPS next day mail. The letter identified the product, the problem, and the action to be taken by the customer. Consignees were instructed to examine their inventory immediately and quarantine any affected product and return to ENDOPATH. If product subject to the recall was further distributed consignees were instructed to notify the other facility and arrange for return of the affected product. Customer were instructed to complete the Business Reply Form confirming receipt of this lette
Quantity213
DistributionWorldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111111ENDOPATH ETS 35MM ENDOSCOPIC LINEAR CUTTER AND RELOADS, AND ETS-FLEX 35MM ENDOSCOPIC ARTICULATING LINEAR CUTTER AND RELO
Ethicon Endo-Surgery, LLC, Jun 8, 2011