Recall Z-2537-2016 ยท DQX Wire, Guide, Catheter

Cook Inc.

Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320, G30490; HPWA-25-260, G30491; HPWA-35-260, G30492; HPWAS-35-260, G30506; HPW-35-150, G56149; HPW-35-180, G56150; HPWS-35-150, G56152; HPWS-35-180, G56153; HPWA-35-80, G56172; HPWA-35-150, G56173; HPWA-35-180, G56174; HPWA-18-180, G56160; HPWAS-35-150, G56176; HPWAS-35-180, G56177. Used with vasc

DateAug 25, 2016 (initiated Jun 24, 2016)
StatusTerminated
ReasonPotential coating contamination with glass particles.
Root causeMaterial/Component Contamination
ActionCook Inc. initiated a voluntary recall of the Roadrunner UniGlide Hydrophilic Wire Guides on June 24, 2016 via courier due to the potential for the device to be contaminated with glass particles. The firm issued a press release dated August 12, 2016. Customers were asked to take the following actions: 1. Examine inventory immediately to determine if you have affected products and quarantine affected products. 2. Return affected products to Stericycle Expert Solutions (a third-party recall administration service provider). Use the enclosed label and a copy of the Acknowledgement and Receip
Quantity8,750 devices were distributed in commerce
DistributionWorldwide distribution. US (nationwide), Australia, Austria, Belgium, Canada, China, Czech Republic, France, Germany, Hong Kong, Hungary, Italy, Jordan, Korea, Spain, South Africa, Taiwan, Turkey, and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110009ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE
Cook, Inc., Feb 2, 2011
K130766ROADRUNNER UNIGLIDE HYDROPHILIC WIRE GUIDE
Cook, Inc., Jul 17, 2013