Recall Z-2540-2016 ยท FTL Mesh, Surgical, Polymeric

TYRX Inc.

TRYX Absorbable Antibacterial Envelope Product Usage: Indicated for stabilization of implanted pacemakers (IPG) and/or implantable cardioverter defibrillators (ICD)

DateJun 29, 2016
StatusTerminated
ReasonTRYX products are being recalled since the processes of spaying, welding, drying oven and polymer were not adequately validated.
Root causeProcess control
ActionMedtronic sent an Urgent Medical Device Recall letter dated June 2016 to affected customers. The letter identified the affected product problem and actions to be taken. Customers are asked to immediately remove and quarantine all unused product that remains in inventory, return unused product to Medtronic and contacted the local Medtronic representative or Customer Service at 800-848-9300 to assist with the return and credit of unused product. Medtronic will provide credit for all non-expired, unused product. The customer should complete the attached Customer Confirmation Certificate and e
Quantity2993 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063091PIVIT CRM (TYRX ANTIMICROBIAL PACEMAKER POUCH), MODEL CMRM-0601
Tyrx Pharma, Inc., Jan 16, 2008