Recall Z-2293-2016 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

MOSAIQ Oncology Information System

DateJul 15, 2016
StatusOpen, Classified
ReasonIt is possible that a change to an Order Set will not be saved in the current open Care Plan. This can result in a wrong chemo dose in the order.
Root causeSoftware Design Change
ActionElekta sent Important Field Safety Notice 371-03-MSQ-001 to all customers on 7/15/2016. The notice informs users of the specific product and version numbers affected by the issue, and any work around that can be used to avoid the issue. The letter also included a response form which is to be returned to Elekta.
Quantity383
DistributionWorldwide distribution. US nationwide, Antigua and Barbuda, Australia, Austria, Bahamas, Bahrain, Canada, Czech Republic, Germany, Hungary, South Africa, Spain, Sweden, Turkey, and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123230MOSAIQ ONCOLOGY INFORMATION SYSTEM
Impac Medical Systems, Inc., Dec 12, 2012