Recall Z-2337-2016 ยท GFD Dermatome
Zimmer Surgical Inc
Zimmer Air Dermatome II Width Plate, 2 inch. Part # 885120200. Width plate is packaged inside of a padded-foam corrugate box. A product label is applied to the lid of the box. The width plates, when sold individually, are placed in a labeled poly zipper bag, which is then placed into a corrugated shipper carton or padded envelope. The Zimmer Air Dermatome II is a skin grafting instrument that i
| Date | Aug 4, 2016 (initiated Jul 4, 2016) |
| Status | Terminated |
| Reason | The surface coating applied to the device is blistering, peeling and discoloring after usage over time |
| Root cause | Process design |
| Action | The firm, Zimmer Biomet Surgical, sent an "URGENT MEDICAL DEVICE RECALL CORRECTION NOTICE" dated 7/6/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Carefully review this letter and ensure all users of the Zimmer Air Dermatome II at your facility have been informed of this notice. 2. Inspect the Zimmer Air Dermatome II devices within your facility and indicate on the attached Certificate of Acknowledgment the condition of the device. Complete and sign the attached Certificate of Acknowledgment and |
| Quantity | 90 units |
| Distribution | US Nationwide Distribution including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.