ClearanceBrief
Recall Z-2335-2016 ยท GFD Dermatome

Zimmer Surgical Inc

Zimmer Air Dermatome II Handpiece w/o Hose, item number 00-8851-001-01. The Zimmer Air Dermatome II handpiece and width plates are packaged inside of a padded-foam corrugate box. An operating manual is packed with the handpiece between the top foam pad and box lid. A product label is applied to the lid of the box. The width plates, when sold individually, are placed in a labeled poly zipper ba

DateAug 4, 2016 (initiated Jul 4, 2016)
StatusTerminated
ReasonThe surface coating applied to the device is blistering, peeling and discoloring after usage over time
Root causeProcess design
ActionThe firm, Zimmer Biomet Surgical, sent an "URGENT MEDICAL DEVICE RECALL CORRECTION NOTICE" dated 7/6/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Carefully review this letter and ensure all users of the Zimmer Air Dermatome II at your facility have been informed of this notice. 2. Inspect the Zimmer Air Dermatome II devices within your facility and indicate on the attached Certificate of Acknowledgment the condition of the device. Complete and sign the attached Certificate of Acknowledgment and
Quantity220 units
DistributionUS Nationwide Distribution including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.