Recall Z-2388-2016 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Edan Diagnostics

EtCO2 Module -Part Number: 02.01.210699 Monitors: iM80, iM50, M50, iM70, iM60, iM8, iM8A, iM8B Patient Monitors and iM60 Vet, iM70 Vet Veterinary Monitors EtCO2 module is an optional component of the following machines: iM80, iM50, M50, iM70, iM60 and iM8, iM8A, iM8B, iM60 Vet, iM70 Vet Monitors, and is intended to be used to measuring expired CO2 as part of multi parameter patient monitors. The

DateJun 23, 2016
StatusTerminated
ReasonIncorrect CO2 readings with some patient monitors.
Root causeNonconforming Material/Component
ActionThe firm, EDAN, sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated 6/23/2016 via email to its customers on 06/23/16. The letter described the product, problem and actions to be taken. Customers were provided two possible solutions. 1. Re-calibrate the affected monitors: a. The customer needs to source the Calibration Gas and other tools locally, EDAN will provide the Specifications of the Calibration Gas; b. EDAN will provide Calibration Instructions to the customer; c. The customer need to calibrate the monitors by themselves and send the Calibration Record back to EDAN. 2. Replace the
Quantity521
DistributionWorldwide Distribution-U.S (nationwide) distribution to the following states; FL, KY, CA; and countries of: Argentina, China, Colombia, , Ethiopia, France, Germany, Guatemala, India, Italy, Lao People's Democratic Republic, Lithuania, Malaysia, Macedonia, Nepal, New Caledonia, Nicaragua, Pakistan, P
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110922PATIENT MONITOR
Edan Instruments, Inc., Jul 20, 2011
K113623PATIENT MONITOR
Edan Instruments, Inc., Feb 1, 2012
K113653PATIENT MONITOR
Edan Instruments, Inc., Feb 6, 2012
K131971IM60/IM70 PATIENT MONITOR
Edan Instruments, Inc., Mar 20, 2014