ClearanceBrief
Recall Z-0056-2017 ยท FMZ Incubator, Neonatal

Natus Medical Incorporated

The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX0006; and KITNC-STLX0007.

DateOct 7, 2016 (initiated Aug 25, 2016)
StatusTerminated
ReasonNatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment.
Root causeUnknown/Undetermined by firm
ActionNatus sent an Urgent Notification Field Safety Notice on August 25, 2016 to all affected customers. Customers are asked to inspect the quality of the welds and wheel alignment for the Natal Care LX Incubators at their facility per attached Inspection Instructions for Assessing the Natal Care LX Wheeled-Based Joint and Weld. Customers are provided with the Natal Care LX Rolling Base Verification Form (DOC-019727) and are requested to verify the weld and gap status for each of the four wheels on the incubator and to note the results on the verification form. After the Verification form is c
Quantity104 units
DistributionUS Distribution to the states of : AL, CA, FL, IL, IN, KY, LA, MA, MS, NJ, NY, OK, PA, SC and VA, and Internationally to Pakistan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070291MEDIX NATAL CARE INTENSIVE CARE INCUBATOR, MODELS-NATAL CARE ST-MX/LX/BX
Medix I.C.S.A., Sep 21, 2007