Recall Z-2669-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Electronics North America Corporation

Allura Xper with release R8.2: Allura Xper FD10, 722026; Allura Xper FD10/10, 722027; Allura Xper FD20, 722028; Allura Xper FD20/10 biplane, 722029; Allura Xper FD10 OR Table, 722033; Allura Xper FD10/10 OR Table, 722034; Allura Xper FD20 OR Table, 722035; Allure Xper FD20/20 , 722038; Allura Xper FD20/20 biplane OR Table, 722039; Allura Xper FD20/15, 722058; Allura Xper FD20/15 OR Table,

DateJul 7, 2016
StatusTerminated
ReasonIncrease in the failure rate of certain low-voltage DC power supplies (DCPS)
Root causeComponent change control
ActionPhilips sent an Urgent - Field Safety Notice dated June 20, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Field Service Engineers will replace the affected DCPS free of charge. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. For further questions regarding this recall please call (978) 659-3000.
Quantity1536 devices (includes both Allura and UNIQ)
DistributionNationwide, Canada, Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Bulgaria CHILE China Colombia Croatia Cuba Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Finland France French Polynesia Georgia Germany Greece Honduras Hong Kong Hungary Icel