Recall Z-2376-2016 ยท FNL Bed, Ac-Powered Adjustable Hospital
Hill-Rom, Inc.
Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this
| Date | Aug 9, 2016 (initiated Jun 14, 2016) |
| Status | Terminated |
| Reason | The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury. |
| Root cause | Device Design |
| Action | Hill-Rom sent an URGENT: MEDICAL DEVICE CORRECTION letter dated June 14, 2016 to the affected US and Canadian customers via email and courier, international customers. The letter included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The letter requests the number of beds in the consignees possession so that repair kits can be provided. If you have any questions regarding this safety notice, contact Hill-Rom Technical Support at 1-800-445-3720, your distributor, your Hill-Rom repres |
| Quantity | 24262 |
| Distribution | Worldwide Distribution - US Nationwide and the countries of: Canada, AUSTRALIA, CHINA, HONG KONG, THAILAND, POLAND, RUSSIAN FEDERATION, SLOVENIA, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, ICELAND, IRELAND, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, SPAIN, SWITZE |
| FDA record | Open on fda.gov |