Recall Z-2587-2016 ยท HJO Biomicroscope, Slit-Lamp, Ac-Powered

Nidek Inc

OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit the values necessary to calculate the power of an IOL for cataract surgery such as axial length, corneal curvature radius, and anterior chamber depth. It measures these necessary values successively through a non-contact optical measurement

DateJul 1, 2016
StatusTerminated
ReasonNidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.
Root causeSoftware in the Use Environment
ActionDear Doctor/Recall letters were sent to affected customers on July 7, 2016. The letters describe the issue and provide instruction should the issue occur. The letters inform users that Nidek will reach out in the next few weeks to schedule installation of new software that will solve the problem. Questions or concerns may be directed to: Customer Service 1.800.722.0219; Monday- Friday 8 am to 5pm PST E-mail: AL-Scan_support@noritsuservice.com.
Quantity17
DistributionUS Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133132OPTICAL BIOMETER AL-SCAN
Nidek Co., Ltd., Jun 11, 2014