ClearanceBrief
Recall Z-2599-2016 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Conmed Corporation

Pediatric Dual Dispersive Electrodes (for patients 5-15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birtcher Systems 6000, 6400, 6500 and CONMED Systems 7500 and 7550 ABC generators or conventional electrosurgery units with similar contact monitoring systems, 10' (3.05m) Cable, Catalog Number 7-383. Intended to be used for the dispersion and return to the electrosurgical generator

DateMay 18, 2016
StatusTerminated
ReasonA version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
Root causeDevice Design
ActionOn May 17, 2016 CONMED Corporation distributed Urgent Device Recall notification letters and Business Reply Forms to their customers via courier service. CONMED Corporation has decided to recall these ThermoGard Dual Dispersive Electrodes for all customers who do not use an electrosurgical generator that employs the Contact Quality Monitoring systems listed. If your electrosurgical generator is not listed, your generator should not be used with the lot codes of the ThermoGard Dual Dispersive Electrodes identified. Customers are advised to inspect inventory for any of the recalled devices.
Quantity4,560 units
DistributionWorldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140658THERMOGARD DISPERSIVE ELECTRODES, THERMOGARD PLUS ABC DISPERSIVE ELECTRODES
Conmed Corporation, Jul 10, 2014