Recall Z-2240-2016 ยท JDO Device, Fixation, Proximal Femoral, Implant

Biomet, Inc.

TSP Hip Fracture Plating System device, fixation, proximal femoral, implant The Biomet Hip Fracture Plating Systems are indicated for use in the open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures and intertrochanteric fractures. Fracture plates with six or more femoral shaft fixation screw holes are also indicated for subtrochanteric

DateJul 19, 2016 (initiated Jun 10, 2016)
StatusTerminated
ReasonThe screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained portion of the screw is below the surface of the plate. A delay in surgery less than 30 minutes may occur.
Root causeDevice Design
ActionZimmer Biomet sent an Urgent Medical Device Recall letter dated June 10, 2016, to all affected customers via courier. Distributors and hospitals were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-6725, Monday through Friday, 8 a.m. to 5 p.m.
Quantity5458 units
DistributionWorldwide Distribution - US (Nationwide) Internationally to Australia, The Netherlands, and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140018TSP HIP FRACTURE PLATING SYSTEM
Biomet, Inc., Apr 4, 2014