Recall Z-2274-2016 ยท JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
Beckman Coulter Inc.
Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic t
| Date | Feb 5, 2016 |
| Status | Terminated |
| Reason | A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration. |
| Root cause | Nonconforming Material/Component |
| Action | Beckman Coulter sent an Important Product Notice Recall letter dated February 5, 2016 to customer. The letter identified the affected product problem and actions to be taken. Customers are instructed to complete and return the enclosed Response Form. Customers with questions are instructed to contact our Customer Support Center " Via website, http://www.beckmancoulter.com/customersupport/support " Via phone, call 800-369-0333 the United States and Canada " Outside the United States and Canada, contact your local Beckman Coulter Representative. |
| Quantity | 62 units total (33 units in US) |
| Distribution | Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.