Recall Z-2351-2016 ยท IFP Formalin, Neutral Buffered

Richard-Allan Scientific Company

Richard Allan Scientific 10% Neutral Buffered Formalin, Part Number: 53901 and 561201 Item 53901 90ml bottle, 125 per case; Item 561201 120ml bottle, 75 per case.

DateJun 13, 2016
StatusTerminated
Reasonpotential contamination with potassium hydroxide may have an impact when performing immunohistochemistry testing. It is unknown whether the contamination will affect tissue antigenicity. If it were to affect tissue antigenicity, the outcome would be risk of false positive or false negative test results which may lead to serious injury. There is the possibility the patient may need to be re-biopsied.
Root causeNonconforming Material/Component
ActionOn 6/13/2016, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected customers via certified mail with instructions for identifying and disposing of the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
Quantity18 cases
DistributionDistributed in the states of FL, PA, NY, MD, CA, and AZ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.