ClearanceBrief
Recall Z-2115-2016 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body.

DateJun 21, 2016
StatusTerminated
ReasonSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
Root causeNonconforming Material/Component
ActionPhilips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.
Quantity1 unit
DistributionDistributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003