Recall Z-0138-2017 ยท FEB Accessories, Cleaning, For Endoscope

Olympus Corporation of the Americas

OER-Pro Endoscope Reprocessor, Endoscope washer/disinfector

DateOct 19, 2016 (initiated Jun 23, 2016)
StatusTerminated
ReasonNew warning requiring Users to conduct duodenoscope precleaning and manual cleaning even when using an Automated Endoscope Reprocessor (AER) that has instructions that may indicate a user could forego certain steps in precleaning and manual cleaning of the endoscopes.
Root causeUnder Investigation by firm
ActionOlympus America Inc. (OAI) mailed a letter to customers informing them that they are issuing an updated Operation Manual for the OER-Pro Automated Endoscope Reprocessor. The following actions were asked to be taken: -Olympus has discontinued previously distributed copies of the OER-Pro Operation Manual. - Inspect your inventory of Operation Manuals and discard any existing inventory of OER-Pro Operation Manuals. - Implement use of the enclosed OER-Pro Operation Manual, which contains the new Warning on pages 73 and 80. - Ensure all reprocessing personnel are completely knowledgeable and
Quantity2, 686 units
DistributionDistributed to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103264ENDOSCOPE REPROCESSOR OER-PRO MODEL OER-PRO
Olympus Medical Systems Corporation, Feb 23, 2011