Recall Z-2529-2016 ยท DOP Mass Spectrometer, Clinical Use

Ab Sciex

AB SCIEX QTRAP 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic

DateAug 12, 2016 (initiated Jun 16, 2016)
StatusTerminated
ReasonMultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.
Root causeSoftware design
ActionAB SCIEX issued recall notification by letter dated June 16, 2016, to all affected customers. The letter identified the problem, provided temporary measures until a permanent software is available. A response form was to be completed and returned. questions regarding this notice contact SCI EX at +1 289 982 2712.
Quantity28 units
DistributionWorldwide Distribution - US (Nationwide) Internationally to Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,