Recall Z-2111-2016 ยท JAK System, X-Ray, Tomography, Computed
Philips Medical Systems (Cleveland) Inc
Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- sectional images of the body.
| Date | Jun 21, 2016 |
| Status | Terminated |
| Reason | Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack and cause injury to patients, operators, by standers or service personnel. |
| Root cause | Nonconforming Material/Component |
| Action | Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt. |
| Quantity | 3 units |
| Distribution | Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K033357 | ACQSIM-MULTISLICE-CT, MODEL 453567042271 Philips Medical Systems (Cleveland), Inc., Jan 9, 2004 |