ClearanceBrief
Recall Z-2109-2016 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Big Bore Oncology tomography X-ray system scanner, model number 728243. The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

DateJun 14, 2016
StatusTerminated
ReasonSocket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.
Root causeNonconforming Material/Component
ActionPhilips Medical Systems (Cleveland), Inc. sent a Customer Information letter dated June 14, 2016, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. No action is required of the customer for this issue. Systems that may have been manufactured with these stainless steel fasteners still meet the safety specifications required for the adjustment cradle and may continue to be used clinically. For further information or support customers were instructed to call 1-800-722-9377. For questions regarding this rec
Quantity47 units
DistributionWorldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004