Recall Z-2718-2016 ยท INQ Table, Powered

ARJOHUNTLEIGH POLSKA Sp. z.o.o.

ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

DateJun 29, 2016
StatusTerminated
ReasonArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees-28 degrees. Although such scenario has been define
Root causeDevice Design
ActionA Field Safety Notice, a Customer Response Form and the Sara Combilizer Addendum to the Instructions For Use (IFU) will be mailed to all affected customers.
Quantity106
DistributionCO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
FDA recordOpen on fda.gov