Recall Z-2147-2016 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Sorin Group USA, Inc.

AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct con

DateJul 8, 2016 (initiated Jun 22, 2016)
StatusTerminated
ReasonSorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.
Root causeProcess control
ActionThe firm, LivaNova, sent an "Urgent Safety Alert" field communication letter dated June 22, 2016 via certified mail to consignees. The letter described the product, problem and actions to be taken. The consignees were instructed to: check your inventory to determine if you own Cannulae, if no inventory, complete the response form and return it per instructions indicated on form.; check the tip angle to determine if it is conforming to specifications, if specifications not met, file a complaint through your normal process whether you decide to keep it, use it or return it to Sorin Group; If you
Quantity5740
DistributionWorldwide Distribution: US (nationwide) to states of: LA, Ml, MO, OH, TN, and TX; and countries of: Italy and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943934AORTIC ARCH CANNULAS
Sorin Group Italia S.R.L., Jun 2, 1995