ClearanceBrief
Recall Z-2242-2016 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Encore Medical, Lp

MODIFIED LINEAR STANDARD/LATERAL NECK TRIALS WITH SPRING SIDES REF S-200635 and S-200636

DateJul 19, 2016 (initiated Jun 21, 2016)
StatusTerminated
ReasonReported failure of the instrument. The retaining ring component of a neck trial became disassembled from the body of the neck.
Root causeProcess design
ActionThe recalling firm notified affected consignees via letter on 6/21/16. The consignees were instructed to return the affected devices.
Quantity14 (7 sets)
DistributionCO, ID, TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974294ENCORE LINEAR POROUS COATED HIP
Encore Orthopedics, Inc., Jan 12, 1998