Recall Z-2188-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Toshiba American Medical Systems Inc

Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System

DateJul 18, 2016 (initiated Feb 24, 2016)
StatusTerminated
ReasonDuring a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
Root causeRadiation Control for Health and Safety Act
ActionToshiba Medical Systems (TAMS) planned action to bring product into Compliance: 1.TAMS will install the software and perform testing to ensure the software update was effective. 2. The customer notification letter which includes a statement that TAMS will, without charge, remedy the defect or bring the product into compliance. . If you have any questions regarding this, please feel free to contact your local Toshiba representative at (800) 521-1968.
Quantity23
DistributionUS Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150967XIDF-AWS801, Angio Workstation, V6.10
Toshibamedical Systems Corporation, Jun 18, 2015