Recall Z-2241-2016 ยท LPS Urinary Homocystine (Nonquantitative) Test System

CSL Behring GmbH

Dimension Vista Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.

DateMay 17, 2016
StatusTerminated
ReasonReagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.
Root causeUnder Investigation by firm
ActionSiemens mailed an Urgent Medical Device Recall to all customers who received the affected lots notifying them of the following: 1) Discontinue use and discard any remaining inventory of the affected lots, and 2) Indicate their replacement product needs within 30 days.
Quantity898 units
DistributionDistributed to: AL, AR, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WI, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063206DIMENSION VISTA HCYS FLEX REAGENT CARTRIDGE; PROTEIN 1 CALIBRATOR; PROTEIN 1 CONTROL LOW, MEDIUM, HIGH