ClearanceBrief
Recall Z-2260-2016 ยท NEP System/Device, Pharmacy Compounding

The Metrix Company

Container fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.

DateJun 23, 2016
StatusTerminated
ReasonAn issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
Root causeDevice Design
ActionOn 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.
Quantity137,598 cases
DistributionProduct distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030888EVA EMPTY SOLUTION CONTAINER
Metrix Co., Apr 2, 2003