ClearanceBrief
Recall Z-2185-2016 ยท HCG Device, Neurovascular Embolization

Stryker Neurovascular

Target Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

DateJul 11, 2016 (initiated Jun 6, 2016)
StatusTerminated
ReasonStryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Root causeProcess change control
ActionStryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instru
Quantity45,302 (8,188 in the U.S.) - total, all model numbers
DistributionWorldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113412TARGET DETACHABLE COIL
Stryker Neurovascular, Dec 13, 2011
K153658Target Detachable Coils
Stryker Neurovascular, May 19, 2016