Recall Z-2131-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

GE Healthcare, Discovery MR750w

DateJun 8, 2016
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.
Root causeNonconforming Material/Component
ActionConsignees were sent on 6/8/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 60901 dated June 8, 2016. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, and Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity471 (184 US; 287 OUS)
DistributionUS: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bermuda Brazil Bulgaria Canada China Egypt France Germany Hungary India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103327DISCOVERY MR750W 3.0T
Ge Healthcare Japan Corporation, Sep 30, 2011
K130115DISCOVERY MR750W 3.0T
Ge Medical Systems, LLC, May 23, 2013
K132376DISCOVERY MR750 3.0T, DISCOVERY MR450 1.5T, DISCOVERY MR750W 3.0T, OPTIMA MR450W 1.5T
Ge Healthcare (Ge Medical Systems, LLC), Nov 15, 2013
K142085DISCOVERY MR750 3.0T ,DISCOVERY MR450 1.5T, DISCOVERY MR750 3.0T,OPTIMA MR450W 1.5T
Ge Medical Systems, LLC, Sep 26, 2014