Recall Z-2118-2016 ยท JAK System, X-Ray, Tomography, Computed
Philips Medical Systems (Cleveland) Inc
Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes..
| Date | Apr 1, 2016 |
| Status | Terminated |
| Reason | Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact. |
| Root cause | Software design (manufacturing process) |
| Action | Philips Medical Systems (Cleveland) Inc. sent an Urgent Field Safety Notice letter to affected customers , via certified letter with receipts returned for tracking purposes. The letter identified the affected product, problem and actions to be taken. Philips Healthcare is implementing software update version 3.5.5 to correct the problem. For questions contact your local Philips representative or local Philips Healthcare office. |
| Quantity | 19 |
| Distribution | US Nationwide Distribution in the states of CA, FL, ID, KS, MI, NY,OH,TX |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K033326 | PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM Philips Medical Systems, Inc., Oct 29, 2003 |