Recall Z-2208-2016 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Sarns TCM; The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradie

DateJul 14, 2016 (initiated Jun 8, 2016)
StatusTerminated
ReasonTerumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Root causeNonconforming Material/Component
ActionTerumo Cardiovascular Systems (Terumo CVS) initiated a voluntary recall on 06/08/2016, by mailing certifed letters to customers. Terumo CVS is notifying all users of its heater-cooler devices that the distributor, Intercon Chemical Company, no longer sells the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions. However, the identical sanitizer is available from other sources. This notice applies to the following Terumo CVS heater-cooler devices: Sarns" TCM and TCM II Cooling and Heating Systems Sarns" Dual Cooler-Heaters HX2" Temperature Managemen
Quantity6412 units
DistributionDomestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K841402TEMPERATURE CONTROL & MONITOR
Sams, Inc., Jun 11, 1984